Anritsu S332C Spezifikationen Seite 7

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 28
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 6
15. Press the VBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
3 kHz. Press ENTER to set
the video bandwidth to 3 kHz.
16. Press the
MARKER key, then the M1 soft
key.
17. Select the
EDIT soft key and use the
Up/down arrow key to center the marker
on the waveform. Verify that the marker
frequency is 1000 MHz, ± 2 kHz for the
S114C, or 2000 MHz ± 4 kHz for the
S332C.
NOTE
If the unit fails the Spectrum Analyzer Fre
-
quency Accuracy test, contact Anritsu
Customer Service.
6. SPECTRUM ANALYZER PHASE NOISE
VERIFICATION
This test can be used to verify the phase noise of the
Site Master Spectrum Analyzer.
a. Equipment Required:
· Anritsu MG3692A Synthesized Signal
Source, with options 2A, 4 and 15A, or
equivalent
b. Procedure:
1. Connect the output of the source to the
Site Master RF Input.
2. Connect the external power supply
(Anritsu part number 40-115) to the Site
Master.
3. Press and hold the
ESCAPE/CLEAR key,
then press the
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
NOTE
Before continuing, allow a 30-minute warm up
for the internal circuitry to stabilize.
4. Set the MG3692A output to 1000 MHz,
with an RF output level of –30 dBm.
5. On the Site Master, press and hold the
ESCAPE/CLEAR key, then press the
ON/OFF key to turn on the Site Master.
(This sets the instrument to the factory
preset state.)
6. Press the
SWEEP key.
7. Press the
RBW soft key and the MANUAL
soft key. Use the Up/down arrow key to
select 10 kHz. Press
ENTER to set the
resolution bandwidth to 10 kHz then
press the
BACK soft key.
8. Press the
VBW soft key and the MANUAL
soft key. Use the Up/down arrow key to
select 3 kHz. Press
ENTER to set the
video bandwidth to 3 kHz.
9. Press the
FREQ/DIST key and the CENTER
soft key.
10. Enter 1000 and press the
ENTER key to
set the center frequency to 1000 MHz.
11. Press the SPAN soft key and enter 0.1.
Press the
ENTER key to set the span to
0.100 MHz.
12. Press the
AMPLITUDE key.
13. Press the
REF LEVEL soft key and enter
27. Press ENTER to set the reference
level to –27 dBm.
14. Press the
MARKER key, then the M1 soft
key.
15. Press
EDIT and enter 1000. Press ENTER
to set the M1 marker frequency to 1000
MHz.
16. Press the BACK soft key and the M2 soft
key.
17. Press
EDIT and enter 1000.03. Press
ENTER to set the M2 marker frequency to
1000.03 MHz (30 kHz higher than the
center frequency).
18. Press the
DELTA (M2-M1) soft key.
19. Press the
RUN/HOLD key and read and
record the D 2 reading
20. Press the RUN/HOLD key to read and re
-
cord five values, then calculate the aver
-
age of the five recorded values.
21. Substract 40 dB from the average value
and verify that the result is £ –75
dBc/Hz. (For example: –35 dBc measured
40 dB = –75 dBc/Hz.)
22. (For example: –35 dBc measured +
(-40 dB) = –75 dBc/Hz.)
Site Master S114C/S332C MM 3