Anritsu S332C Spezifikationen Seite 11

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 28
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 10
9. Press the FREQ/DIST key and the CENTER
soft key.
10. Enter
1 and press the GHz soft key to set
the center frequency to 1GHz.
NOTE
The Measured Occupied BW value for each
RBW setting can be recorded in Table 1.
1 MHz RBW Test
11. Press the
SPAN soft key, enter 5 and press
the
MHz soft key to set the span to 5
MHz.
12. Press the
SWEEP key.
13. Press the
RBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
1MHz. Press ENTER to set
the resolution bandwidth to 1 MHz and
select
BACK.
14. Press the
VBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
3 kHz and press ENTER to set
the video bandwidth to 3 kHz.
15. Press the
MEASURE soft key, the OBW soft
key and then the
METHOD soft key to
measurea3dBbandwidth.
16. Press the
MEASURE soft key and read the
displayed Meas Occ BW value. The Meas
Occ BW value should be ± 20% of the
RBW.
100 kHz RBW Test
17. Press the
FREQ/DIST key.
18. Press the
SPAN soft key, enter 1 and press
the
MHz soft key to set the span to 1
MHz.
19. Press the
SWEEP key.
20. Press the
RBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
100 kHz. Press ENTER to set
the resolution bandwidth to 100 kHz and
select
BACK.
21. Press the
MEASURE soft key, the OBW soft
key and then the
METHOD soft key to
measurea3dBbandwidth.
22. Press the
MEASURE soft key and read the
Meas Occ BW value. The Meas Occ BW
value should be ± 20% of the RBW.
30 kHz RBW Test
23. Press the
FREQ/DIST key.
24. Press the
SPAN soft key, enter .1 and
press the
MHz soft key to set the span to
0.1 MHz.
25. Press the
SWEEP key.
26. Press the
RBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
30 kHz. Press ENTER to set
the resolution bandwidth to 30 kHz and
select
BACK.
27. Press the
MEASURE soft key, the OBW
soft key and then the METHOD soft key to
measurea3dBbandwidth.
28. Press the
MEASURE soft key and read the
displayed Meas Occ BW value. The Meas
Occ BW value should be ± 20% of the
RBW.
10 kHz RBW Test
29. Press the
FREQ/DIST
30. Press the SPAN soft key, enter .04 and
press the
MHz soft key to set the span to
0.04 MHz.
31. Press the SWEEP key.
32. Press the
RBW soft key and then the EDIT
soft key. Use the Up/down arrow key to
select
10 kHz. Press ENTER to set the reso-
lution bandwidth to 10 kHz and select
BACK.
33. Press the
MEASURE soft key, the OBW
soft key and then the METHOD soft key to
measurea3dBbandwidth.
34. Press the
MEASURE soft key and read the
displayed Meas Occ BW value. The Meas
Occ BW value should be ± 35% of the
RBW.
Site Master S114C/S332C MM 7
RBW
Setting
Lower
Limit
Measured
Value
Upper
Limit
1 MHz 0.8 MHz 1.2 MHz
100 kHz 80 kHz 120 kHz
30 kHz 24 kHz 36 kHz
10 kHz 6.5 kHz 13.5 kHz
Table 1. RBW Check
Seitenansicht 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 27 28

Kommentare zu diesen Handbüchern