10. SPECTRUM ANALYZER MEASUREMENT
ACCURACY TEST
This test verifies the level accuracy of the Site Mas
-
ter Spectrum Analyzer.
a. Equipment Required:
·
Anritsu MG3692A Synthesized Signal
Source, with options 2A, 4 and 15A
·
Anritsu ML2430A-Series Power Meter or
equivalent
·
Anritsu MA2442A High Accuracy Power
Sensor or equivalent
·
Anritsu N241A50 Power Splitter or
equivalent
·
Anritsu 34NN50A 50 ohm adapter or
equivalent
·
Anritsu 34RSN50 50 ohm adapter or
equivalent
·
Anritsu 15NNF50-1.5C RF Coaxial Cable
or equivalent
b. Procedure:
1. Connect the power sensor to the power
meter and calibrate the sensor.
2. Using the power splitter, coaxial cable
and adapters, connect the Site Master to
the signal source and the power sensor
as shown in Figure 1.
3. Connect the external power supply
(Anritsu part number 40-115) to the Site
Master.
4. On the Site Master, press and hold the
ESCAPE/CLEAR key, then press the
ON/OFF key to turn on the Site Master.
(This sets the instrument to the factory
preset state.)
5. Turn on the power meter and signal
source.
NOTE
Before continuing, allow a 30-minute warm up
for the internal circuitry to stabilize.
6. On the Site Master, press the
MODE key.
Use the up/down arrow key to highlight
SPECTRUM ANALZYER and press ENTER
to select spectrum analyzer mode.
7. Press the
SWEEP key.
8. Press the
RBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
10 kHz. Press ENTER to set
the resolution bandwidth to 10 kHz.
9. Press
BACK, then the VBW soft key and
then the
MANUAL soft key. Use the
Up/down arrow key to select
3 kHz. Press
ENTER to set the video bandwidth to 3
kHz.
10. Press
BACK, then MEASURE, then DETEC-
TION
, then AVERAGE.
11. Press the
FREQ/DIST key.
8 Site Master S114C/S332C MM
M o d u l a t io n
L i n e
O p e ra te
S ta n d b y
L e v e l
F re q u e n c y
S y s t e m
O u t p u t
O n O ff
R F O u t p u t
5 0
9
C le a r
E n t r y
B a c k
S p a c e
7 8 9
4 5 6
1 2 3
0
.
+ / -
MA2442A
SENSOR
ML2438A POWER METER
MG3692A SOURCE
S332C
15NNF50-1.5C
34NN50A
ADAPTER
34RSN50 ADAPTER
N241A50
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT
MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
MODE
FREQ/DIST
AMPLITUDE
SWEEP
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
/
1
2
4
5
6
7
8
9
0
3
+
-
.
Site Master S332C
Figure 1. Measurement Accuracy Setup
Kommentare zu diesen Handbüchern