19. Press the FREQ/DIST key and the START
soft key.
20. Enter
4 and press the MHz soft key to set
the start frequency to 4 MHz.
21. Press the
STOP soft key and enter 1000,
then press the
MHz soft key to set the
stop frequency to 1000 MHz.
22. Wait until one full sweep is complete.
23. Press the
MARKER key and then the M1
soft key.
24. Press the
ON/OFF soft key and then the
MARKER TO PEAK soft key.
25. Record the M1 amplitude reading and
verify it is £ –90 dBm.
26. Press the
FREQ/DIST key and the START
soft key.
27. Enter
1000 and press the MHz soft key to
set the start frequency to 1000 MHz.
28. Press the
STOP soft key.
29. For the S114C, enter
1600, then press the
MHz soft key to set the stop frequency to
1600 MHz.
For the S332C, enter
2000, then press
MHz soft key to set the stop frequency to
2000 MHz.
30. Wait until one full sweep is complete.
31. Press the
MARKER key and then the M1
soft key.
32. Press the
ON/OFF soft key and then the
MARKER TO PEAK soft key.
33. Record the M1 amplitude reading and
verify it is £ –90 dBm.
NOTE
Continue on to Step 34 only if the unit being
tested is a Model S332C Site Master.
34. Press the
FREQ/DIST key and the START
soft key.
35. Enter
2000 and press the MHz soft key to
set the start frequency to 2000 MHz.
36. Press the
STOP soft key and enter 3000,
then press the
MHz soft key to set the
stop frequency to 3000 MHz.
37. Wait until one full sweep is complete.
38. Press the
MARKER key and then the M1
soft key.
39. Press the
ON/OFF soft key and then the
MARKER TO PEAK soft key.
40. Record the M1 amplitude reading and
verify it is £ –90 dBm.
9. SPECTRUM ANALYZER RESOLUTION
BANDWIDTH ACCURACY TEST
The following test can be used to verify the resolu
-
tion bandwidth accuracy of the Site Master Spec
-
trum Analyzer at four frequencies.
a. Equipment Required:
·
Anritsu MG3692A Synthesized Signal
Source with options 2A, 4 and 15, or
equivalent
·
10 MHz Reference Standard
·
3 GHz RF cable (Nm to Nm) or equiva
-
lent
b. Procedure:
1. Connect the 10 MHz reference source to
the Anritsu MG3692A Synthesized Sig-
nal Source.
2. Connect the output of the Anritsu
MG3692A Synthesized Signal Source to
the Site Master Spectrum Analyzer RF
Input.
3. On the Site Master, press and hold the
ESCAPE/CLEAR key, then press the
ON/OFF key to turn on the Site Master.
(This sets the instrument to the factory
preset state.)
NOTE
Before continuing, allow a 30-minute warm up
for the internal circuitry to stabilize.
4. Set the MG3692A output to 1 GHz, with
an RF output level of –30 dBm.
5. On the Site Master, press the
MODE key.
Use the up/down arrow key to highlight
SPECTRUM ANALZYER and press ENTER
to select spectrum analyzer mode.
6. Press the
AMPLITUDE key and the REF
LEVEL
soft key.
7. Enter –27 and press the
ENTER key to set
the Reference Level to –27 dBm.
8. Press the SCALE soft key and enter 3,
then press ENTER.
6 Site Master S114C/S332C MM