11. DISPLAYED AVERAGE NOISE LEVEL
The following test can be used to verify the displayed
average noise level of the Site Master Spectrum Ana
-
lyzer.
a. Equipment Required:
·
Anritsu 28N50-2 or SM/PL 50 ohm Ter
-
mination or equivalent
b. Procedure:
1. Connect the 50 ohm termination to the
Site Master Spectrum Analyzer RF In
-
put.
2. On the Site Master, press and hold the
ESCAPE/CLEAR key, then press the
ON/OFF key to turn on the Site Master.
(This sets the instrument to the factory
preset state.)
NOTE
Before continuing, allow a 30-minute warm up
for the internal circuitry to stabilize.
3. On the Site Master, press the
MODE key.
Use the up/down arrow key to highlight
SPECTRUM ANALZYER and press ENTER
to select spectrum analyzer mode.
4. Press the
AMPLITUDE key and the REF
LEVEL
soft key.
5. Enter –
75 and press the ENTER key to set
the Reference Level to –75 dBm.
6. Press the
SCALE soft key and enter 5,
then press ENTER.
7. Press the
SWEEP key and then the MEA
-
SURE
soft key.
8. Press the
DETECTION soft key, then press
AVERAGE soft key.
9. Press the
SWEEP key.
10. Press the
RBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
10 kHz. Press ENTER to set
the resolution bandwidth to 10 kHz and
select
BACK.
11. Press the
VBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
100 Hz and press ENTER to
set the video bandwidth to 100 Hz.
12. Press the
FREQ/DIST key and the START
soft key.
13. Enter
100 and press the kHz soft key to
set the start frequency to 100 kHz.
14. Press the STOP soft key and enter 500,
then press the
kHz soft key to set the stop
frequency to 500 kHz.
15. Wait until a full sweep is complete.
16. Press the
MARKER key and then press
the
M1 soft key.
17. Press the
MARKER TO PEAK soft key. Re
-
cord the M1 amplitude reading and ver
-
ify it is £ –80 dBm.
18. Press the
FREQ/DIST key and the START
soft key.
19. Enter
500 and press the kHz soft key to
set the start frequency to 500 kHz.
20. Press the
STOP soft key and enter 4, then
press the
MHz soft key to set the stop fre
-
quency to 4 MHz.
21. Wait until a full sweep is complete.
22. Press the
MARKER key and then press
the
M1 soft key.
23. Press the
MARKER TO PEAK soft key. Re-
cord the M1 amplitude reading and ver-
ify it is £ –95 dBm.
24. Press the
SWEEP key.
25. Press the
VBW soft key, and then the
MANUAL soft key. Use the Up/down arrow
key to select
3 kHz, and press ENTER to
set the video bandwidth to 3 kHz.
26. Press the
FREQ/DIST key and the START
soft key.
27. Enter
4 and press the MHz soft key to set
the start frequency to 4 MHz.
28. Press the
STOP soft key and enter 1000,
then press the MHz soft key to set the
stop frequency to 1000 MHz.
29. Wait until a full sweep is complete.
30. Press the
MARKER key and then press
the
M1 soft key.
31. Press the
MARKER TO PEAK soft key. Re
-
cord the M1 amplitude reading and ver
-
ify it is £ –95 dBm.
32. Press the
FREQ/DIST key and the START
soft key.
33. Enter
1000 and press the MHz soft key to
set the start frequency to 1000 MHz.
10 Site Master S114C/S332C MM