16. Adjust the signal levels of both signal
sources so that both signal peaks are at
–20 dBm as indicated on the Site Master
Spectrum Analyzer display.
17. Referring to steps 13 through 15, set the
RBW to 10 kHz and the VBW to 1 kHz.
18. Wait until the sweep is complete, then
reduce the power level of the second
source signal to below –85 dBm and ver
-
ify that the signal is visible at below –85
dBm.
NOTE
Be sure to wait until the sweep is complete be
-
fore changing the power level of the second
source.
13. POWER MONITOR VERIFICATION
If the Power Monitor (Option 5) is installed in the
Site Master, the following test can be used to verify
the accuracy of the power measurements. Measure
-
ment calibration of the Site Master is not required
for this test.
a. Equipment Required:
· RF Detector, 10 MHz to 20 GHz,
Anritsu 560-7N50B
· 10 dB Attenuator, Weinschel 1-10
· 30 dB Attenuator, Weinschel 1-30
· RF Reference Source, 0.050 GHz,
Anritsu MA2418A
·
DC Power Supply, Anritsu 2000-933
b. Procedure
1. Connect the DC power supply to the
MA2418A Reference Source (Figure 2).
2. Connect the MA2418A Reference Source
to the input of the 560-7N50B RF detec
-
tor.
3. Connect the RF Detector output to the
RF Detector input of the Site Master.
4. Connect the DC power supply to the ap
-
propriate line voltage to supply power to
the MA2418A Reference Source.
5. Press and hold the
ESCAPE/CLEAR key,
then press the
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
6. Press the
MODE soft key.
7. Use the Up/Down arrow key to highlight
POWER MONITOR, then press ENTER.
8. Press the
ZERO soft key to zero the power
monitor. When complete,
ZERO ADJ:ON is
displayed in the message area.
9. Verify that the power monitor reading is
0.0 dBm ± 1dB.
10. Connect the output of the MA2418A Ref-
erence Source to the two attenuators so
as to add 40 dB of attenuation (Figure 2).
11. Connect the MA2418A Reference Source
and the attenuators to the input of the
560-7N50B RF detector.
12. Verify that the power monitor reading is
now –40.0 dBm ±2dB.
12 Site Master S114C/S332C MM
DC Power Supply 2000-933
Attenuators
Site Master w/ Option 5 Installed
MA2418A
Reference Source
560-7N50B
RF Detector
15-24V DC
20 mA
MA2418A Reference Source
50 MHz, 50
0dBm
W
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT
MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
MODE
FREQ/DIST
AMPLITUDE
SWEEP
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
/
1
2
4
5
6
7
8
9
0
3
+
-
.
Site Master S331B
Figure 2. Power Monitor Verification