12. Press the SPAN soft key and enter .5,
then press the
MHz key to set the span to
.5 MHz.
13. Press the
AMPLITUDE key.
14. Press the
REF LEVEL soft key, enter 20
and press the ENTER key to set the refer
-
ence level to +20 dBm.
15. Press the
FREQ/DIST key and the CENTER
soft key.
16. Enter
50 and press the MHz soft key to
set the center frequency to 50 MHz.
17. On the Anritsu ML2430A-Series Power
Meter, press the
Sensor key and then the
CalFactor soft key. Select the FREQ soft
key and enter 50 MHz for the Input Sig
-
nal Frequency. This sets the power meter
to the proper power sensor cal factor.
18. Set the MG3692A output to 50 MHz CW
and the power level to +10 dBm as indi
-
cated on the power meter.
NOTE
To insure accuracy, set the output of the signal
source using the power meter reading. Do not
rely on the signal source display.
19. Press the
MARKER key, then the M1 soft
key.
20. Select the MARKER TO PEAK soft key to
position the marker at the center of the
response for the test frequency.
NOTE
Table 2 can be used to record the results of
this test for future reference.
21. Verify that the M1 reading is ± 2dB
maximum from the input signal.
22. Press the
AMPLITUDE key.
23. Press the
REF LEVEL soft key, enter 0 and
press
ENTER to set the reference level to
0 dBm.
24. Set the MG3692A power level to –10
dBm as indicated on the power meter.
25. Verify that the M1 reading is ± 2dB
maximum from the input signal.
26. Press the
AMPLITUDE key.
27. Press the
REF LEVEL soft key, enter 20
and press ENTER to set the reference
level to –20 dBm.
28. Set the MG3692A power level to –30
dBm as indicated on the power meter.
29. Verify that the M1 reading is ± 2dB
maximum from the input signal.
30. Press the
AMPLITUDE key.
31. Press the
REF LEVEL soft key and enter
–40. Press
ENTER to set the reference
level to –40 dBm.
32. Set the MG3692A power level to –50
dBm as indicated on the power meter.
33. Verify that the M1 reading is ± 2dB
maximum from the input signal.
34. Repeat steps 15 through 33 for frequen
-
cies of 500 MHz, 1000 MHz, 1500 MHz,
2000 MHz, 2500 MHz and 2950 MHz.
Site Master S114C/S332C MM 9
Freq (MHz)
Power
Level
(dBm)
Site Master
Ref Level
(dBm)
M1 Reading
50 +10
–10
–30
–50
+20
+20
–20
–40
500 +10
–10
–30
–50
+20
+20
–20
–40
1000 +10
–10
–30
–50
+20
+20
–20
–40
1500 +10
–10
–30
–50
+20
+20
–20
–40
2000
(S332C only) +10
–10
–30
–50
+20
+20
–20
–40
2500
(S332C only) +10
–10
–30
–50
+20
+20
–20
–40
2950
(S332C only) +10
–10
–30
–50
+20
+20
–20
–40
Table 2. Measurement Accuracy Test Results