Step 17. Press the EDIT soft key to turn Marker M2 ON and use Up/down arrow key to
place Marker M2 at the center of the second harmonic. Press ENTER to set the
marker position.
Step 18. Press the
DELTA (M2-M1) soft key. This will provide frequency and amplitude
information with respect to marker M1.
Step 19.
Press the MARKER key and select the
M3 soft key.
Step 20. Press the
EDIT soft key to turn Marker M3 ON and use the Up/down arrow key
to place Marker M3 at the center of the third harmonic.
Step 21. Press the
DELTA (M3-M1) soft key. This will provide frequency and amplitude
information with respect to marker M1.
Figure 5-14 shows the distortion characteristics of the signal source just measured. Marker
1 indicates the fundamental frequency. Marker 2 shows that the second order distortion of
the source is 46.24 dB below the carrier. Marker 3 shows that the third order harmonic is
48.44 dB below the fundamental frequency.
5-22
Chapter 5 Spectrum Analyzer Measurements
Figure 5-14. Harmonic Distortion