Step 8.
Press the MEASURE key and press the
OBW soft key. Select the measurement
method (dB Down or % of Power) by pressing the
METHOD soft key. The cur-
rently selected method is shown in the lower left of the display.
Step 9. Press the dBc or % soft keys to adjust the settings as appropriate.
Step 10. Press the
MEASURE ON/OFF soft key to initiate the measurement. OBW will
appear to the left of the graph when the occupied bandwidth measurement is on.
Figure 5-6 shows the occupied bandwidth results using the % of power method on a CDMA
signal.
When the occupied bandwidth measurement is on, an
OBW icon appears to the left of the
display. Occupied bandwidth is calculated at the end of a sweep. An hourglass is displayed
as the calculations are performed.
5-11
Chapter 5 Spectrum Analyzer Measurements
Figure 5-6. Occupied Bandwidth
CW IS-95 CDMA
Reference Level –15 dBm –10 dBm
Input Attenuation Level Auto Auto
Resolution Bandwidth Auto 10 kHz
Video Bandwidth Auto 30 kHz