Seitenansicht 162
Related
Commands:
SYSLD
TEXT User text command
SYSTEM
Syntax:
TEXT <text string>
text string:
Text string of up to 20 characters
Remarks:
Defines the text string that will be displayed using the TEXTS command.
Related
Commands:
TEXTS
Query:
TEXT?
Returned
String:
TEXT <text string>
TEXTS User text display command
SYSTEM
Syntax:
TEXTS <state>
state:
ON or OFF
Remarks:
This command turns on or off the display of text entered using the TEXT
command. Up to 20 characters of user text can be displayed on the top line
of the data screen for READOUT, PROFILE and PWRvsTIME display
modes.
Related
Commands:
TEXT
Query:
TEXTS?
Returned
String:
TEXTS <state>
TR0 Trigger hold mode
GPIB TRIGGER
Syntax:
TR0
Remarks:
Sets both channels to trigger hold mode. It does not trigger until it receives
a TR1 or TR2 or GET (group executive trigger), *TRG or TR3 command. If
it receives a TR3 command it reverts back to the trigger mode it was in be
-
fore the TR0 command was sent. If the REM line is low, this command has
no effect.