Remarks:
The LLIMS command turns the limits on and off.
Related
Commands:
LLIM
Query:
LLIMS? <c>
Returned
String:
LLIMS <c>,<state>
MMRST Min Max Tracking reset
CHANNEL
Syntax:
MMRST <c>
c:
1 or 2
Remarks:
This command resets the min/max values when in 'Readout' or 'Power vs..
Time' mode. In profile mode, this command is used to reset the channels
min/max values.
MNGDB Output Min Graph Binary Data
DATA OUTPUT
Syntax:
MNGDB
Remarks:
Available in graph modes only. Outputs in binary form the min graph data
to the GPIB in the long integer form of 1024 bits per dB as a definite length
arbitrary block response data. The C programming example 'Binary output
decoding' on page 6-133 shows how to extract the binary data. The re
-
sponse form is as follows :
MNGDB <#><length><number_of_bytes><data_byte_1><data_byte_2>
...<data_byte_n><\n>
<length> number of ASCII characters make up the number_of_bytes value
<number_of_bytes> number of bytes of data contained in rest of the string
<data_byte_n> four of these values makes up the long integer.
For example: FF FF D1 64 = –11932 As it is based on 1024 per dB, divide
by 1024 to get the dB value (–11.652).
MNGD Output Min Graph Data
DATA OUTPUT
Syntax:
MNGD
Remarks:
Available in graph modes only. Outputs in ASCII form the min graph data.
The format is as follows:
ML2430A OM 6-53
GPIB OPERATION ML24XXA NATIVE COMMANDS