port:
1 or 2
val:
–5.00 to +5.00 Volts
Remarks:
Sets up the stop value for the voltage output in analog output mode. At
-
tempting to set the start value to a voltage greater than the stop value, or
the stop value lower than the start value, will result in an execution error.
Query:
OBVSP? <port>
Returned
String:
OBVSP <port>,<val>
OBVST BNC analog output start voltage scale
BNC
Syntax:
OBVST <port>,<val>
port:
1 or 2
val:
–5.00 to +5.00 Volts
Remarks:
Sets up the start value for the voltage output in analog output mode. At-
tempting to set the start value to a voltage greater than the stop value, or
the stop value lower than the start value, will result in an execution error.
Query:
OBVST? <port>
Returned
String:
OBVST <port>,<val>
OBZL BNC RF blanking output level when zeroing
BNC
Syntax:
OBZL <level>
level:
HIGH (TTL high)
LOW (TTL low)
Remarks:
Sets the TTL level of the BNC RF blanking output.
Query:
OBZL?
Returned
String:
OBZL <level>
OFFCLR Clear an offset table
SENSOR
Syntax:
OFFCLR <val>
val:
1 to 5
Remarks:
Sets all the values in the table specified to 0 dB and 0.00 Hz.
6-60 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION