DIGITAL LINK MEASURING INSTRUMENTS
178 For product ordering information, see pages 5 – 12
Wander generation: 10 µHz to 10 Hz, 0 to 400,000 UIp-p (1 UIp-p steps), conform to ITU-T O.172
Measurement functions: UIp-p, UI + peak, UI – peak, UIrms, hit count, hit second, %F second, peak jitter
Measurement mode: Repeat, single, manual
Display: Current, last
Measurement interval: 1 to 99 s, 1 to 99 min, 1 to 99 h, 1 to 99 day
Jitter measurement: Conform to ITU-T O.172
UIp-p measurement
2 UI range: 0.000 to 2.020 UIp-p (0.001 UIp-p steps)
20 UI range: 0.00 to 20.20 UIp-p (0.01 UIp-p steps)
1000 UI range: 0 to 1010 UIp-p (1 UIp-p steps, 2488.32 Mbit/s only)
4000 UI range: 0 to 4040 UIp-p (2 UIp-p steps, 9953.28 Mbit/s only)
UI rms measurement
2 UI range: 0.000 to 0.714 UIrms (0.001 UIrms steps)
20 UI range: 0.00 to 7.17 UIrms (0.01 UIrms steps)
Filters:
Confirming to ITU-T O.172 and Bellcore GR1377
LP, HP0 + LP, HP1 + LP, HP1' + LP, HP2 + LP, HP + LP, HP' + LP, LP' (1000/4000 UI range only),
HP0 + LP' (1000/4000 UI range only)
Bit rate F0 F1 F2 A0 A1 Steps
(bit/s) (µHz) (mHz) (Hz) (UIp-p) (UIp-p) (UIp-p)
2488.32M 10 400 10 400,000 16,000 1
9953.28M 10 400 10 400,000 16,000 1
Bit rate Range F1 F1' F2 F2' F3 F4
(bit/s) (UI) (Hz) (Hz) (kHz) (kHz) (MHz) (MHz)
2488.32M
2 — 100 — 100 1 20
20 10 — 10 — 120
9953.28M
2 — 100 — 400 4 80
20 10 — 40 — 480
Bit rate Range F1 F1' F2 F2' F3
(bit/s) (UI) (Hz) (Hz) (Hz) (Hz) (kHz)
2488.32M 1000 — 1 — 12.1 5
9953.28M 4000 1 — 12.1 — 20
Bit rate HP0 HP1 HP1' HP2 HP' HP LP LP'
(bit/s) (Hz) (kHz) (kHz) (MHz) (kHz) (kHz) (MHz) (kHz)
2488.32M 10 5 — 1 — 12 20 5
9953.28M 10 10 20 4 50 12 80 20
Wander generation
Jitter measurement