Anritsu ME7840A Bedienungsanleitung Seite 46

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 126
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 45
sure that the equipment is properly connected to the bus. If
so, the following screen will be displayed:
Step 3. Click on “Run Calibration Procedure” to display a list of cali
-
bration specification files (below). (Refer to paragraph 4-3 for
a description of these files.) Running the files is described in
succeeding numbered paragraphs.
4-6 POWER LEVEL The A_Power Level.txt specification file provides prompts for a power
level calibration. Proceed as follows:
Step 1. Zero-set the power sensor for accurate measurement in fol
-
lowing steps.
Step 2. Select “Tools” then “Run Calibration Procedure.”
POWER LEVEL OPERATIONS, CALIBRATION
4-4 ME 7840A OMM
Seitenansicht 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 125 126

Kommentare zu diesen Handbüchern

Keine Kommentare